1 installing the monitor, 1 operating environment, 2 the gas sample line – LumaSense Technologies INNOVA 1316A-1 User Manual

Page 18

Advertising
background image

Chapter 2
_________________________________________________________________________________________

_____________________________________________________________________________

BE6020-14

1316A-1 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 18 of 98

2.1 Installing the Monitor

The rear panel has a serial-link and a USB socket for the PC

interface. The mains input, sample gas inlet and outlet for expelled

gas samples are located on the rear panel.

2.1.1 Operating Environment

The following environmental conditions must be observed:

Safe ambient operating temperatures range from 5°C to 40°C

Humidity should not exceed 90% RH non-condensing.

Warning! The monitor is not intended for use in explosive

environments.

2.1.2 The Gas Sample Line

The gas to be sampled (measured) is drawn from the sample point

via a sample line at a flow rate of 800 ml/min. Various types of

sample lines are available depending on the individual application.

The sample line tubing between the sample point and the monitor

should be kept as short as possible to assist response times. If you

are unsure about the type of sample line for your application, contact

your LumaSense representative for advice before installing the

system.

Up to 100 m tube (AF0614) with internal diameter 3mm can be

attached. The response time is then 70 sec.

Note: You must use a filter in the gas-line immediately before the

gas-line enters the monitor. The UD 5091A in-line filter (optional

accessory) is ideal for this purpose.

The gas inlet for the sample line is located on the rear panel of the

monitor. If the dew point of the gas sample is comparable to the

temperature in the monitor, Nafion tubing UD5037 can be used to

remove water and eliminate the risk of condensation inside the

monitor

Nafion is impermeable to O

2

, CO

2

, CO and most hydrocarbons,

whereas alcohols and other gases will completely or partly penetrate

through the material. Contact an LumaSense Technical Support

representative for advice on whether or not to use Nafion when

monitoring a particular gas.

You should protect the Nafion tubing against the effects of ultra-

violet light, as this reduces the tubes lifetime. Nafion tubing should

Advertising