Figure 4.33: Data Status window to select Hydro Carbon Mode.

4.3.7 To perform a span calibration with multi component gas mixture.

If more than one gas is to be span calibrated at the same time this

can be performed using multi component gas mixtures.

In the Span calibration window (see

figure 4.22

) select the gases to

be calibrated and key in the appropriate concentrations from the

certificate from the gas cylinder.

Before doing the Span calibration please make sure that Zero

calibration is set to “Manual” and the mode is set to 3 as described in

section 4.3.5

.

1. Use to tick which of the CO, CO

2

, CHx gasses that you want

to be calibrated.

2. Type in the correct concentrations for the ticked CO, CO

2

, CHx

gasses.

3. Connect the calibration-gas bottle as described in

section

4.3.4

.

4. Open the gas valve and adjust the flow so that a small surplus

flow exists in the side-stream tube.