8 installing a new no sensor – LumaSense Technologies INNOVA 1316A-1 User Manual

Page 85

Advertising
background image

Chapter 4
_________________________________________________________________________________________

_____________________________________________________________________________

BE6020-14

1316A-1 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 85 of 98

After making the gain Calibration please make sure that the Mode is set

back into 5 and that the Zero calibration is set to “Auto”, as described in

section 4.3.5

.

4.3.8 Installing a new NO sensor.


After installation the new NO sensor it must be span calibrated following

the procedure described in the previous sections.
After performing the NO span calibration:
Select the New NO sensor in the Bench command menu, see figure

4.35.

Figure 4.35. Bench command menu select New NO sensor


The following window appears.