Chapter 4
_________________________________________________________________________________________

_____________________________________________________________________________

BE6020-14

1316A-1 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 86 of 98

Figure 4.36 . The New NO sensor window.


If a new sensor is installed Press SEND.
Warning: never attempt to send the new NO sensor command after a

NO span calibration unless a new NO sensor has been installed.

The Low NO Sensor Limit is calculated and saved.

4.4 Reset Span Calibration


At any time it is possible to return to the factory saved calibration

factors.
Pull down the Bench Command Menu.