Press the Send Button and the Instrument is Reset and ready for set-

up.

3.6 Operating Modes of the Multi Gas Monitor

1316A-1

The Multi Gas Monitor 1316A-1’s operations comprise four modes of

operation.

Start-up

Normal

Standby

System Fault

3.6.1 Start-up Operating Mode

After POR (power on / reset), the Multi Gas Monitor 1316A-1

performs the following:

Start-up and initialization.

Self-test.

Thermal warm-up. After a cold start or a Software Reset

bench command the warm-up time period is 35 seconds

maximum. After leaving standby mode the typical warm-

up time period is 20 seconds after completion of the first

zero.

The following actions are not allowed during the start-up mode:

Zero calibration.

Span calibration.

3.6.2 Normal Operating Mode

When the Start-up time has elapsed, the Multi Gas Monitor 1316A-1

enters Normal Operating Mode.

All actions are allowed, like measurement, calibration or any other

bench command.

If measurement has not been requested for >2 minutes, the Multi

Gas Monitor 1316A-1 leaves Normal mode and enters Standby

mode.

3.6.3 Standby Operating Mode

The 1316A-1 Gas Monitor’s Pump is turned OFF when the Standby

mode is entered.