2 error and warnings related to a zero calibration, Figure 4.5 – LumaSense Technologies INNOVA 1316A-1 User Manual

Page 61

Advertising
background image

Chapter 4
_________________________________________________________________________________________

_____________________________________________________________________________

BE6020-14

1316A-1 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 61 of 98

The Proc In Progress in

figure 4.3

is changed from a red In

Progress to a green Not in Progress in the status screen, see

figure

4.5

.

Figure 4.5. Status Window showing Zero not required.


The 30-minute zero request interval timer is reset.

4.2.2 Error and Warnings Related to a Zero Calibration

The following conditions cause the Multi Gas Monitor –INNOVA 1316A-

1 to refuse to execute the zero calibration.

1.

The Multi Gas Monitor –INNOVA 1316A-1 is in Start-Up mode,

see

figure 4.6

, Operating Mode.

Advertising