ITC IRMA TruPoint Blood Analysis System User Manual

Page 101

Advertising
background image

7.5

4. If BEb was selected, chose the source of the patient hemoglobin value to

be used in the BEb calculation and press

next

. Highlight the button next

to one of the following options:

a. Default value: 15g/dL is automatically

used in all BEb calculations.

b. Manually Entered value: User manually

enters a patient hemoglobin value with each
test.

c. Calculated value: If a CC cartridge is run with both blood gas and hct

analytes selected, the tHb value that is calculated from the hct test will
be used in the BEb calculation.

If Default or Manually Entered was selected, Calculated Values setup is
complete.

5. If

Calculated was selected, chose an alternate tHb source to be used

when a calculated tHb is not available, and press

next

. Highlight the

button next to one of the following options:

a. Default value: 15 g/dL is automatically

used in the BEb calculation.

b. Manually Entered value: User manually

enters a patient hemoglobin value with each
test.

Calculated Values setup is complete.

Selecting the formulas to be used to derive the calculated values

Selecting Base Excess and Bicarbonate Formulas:

1. Highlight

Formulas

and press

edit

.

2. Highlight the button next to the formula to be

used in the calculation of base excess and
bicarbonate. Press

next

.

Advertising