ITC IRMA TruPoint Blood Analysis System User Manual

Page 45

Advertising
background image

2.15

Patient Temperature

• The

Patient Temperature option allows entry of

a patient temperature when performing a blood
gas test. A default temperature of 37°C auto-
matically displays if a patient ID is not entered,

or no previous blood gas results exist for a patient ID. Press

next

to accept

the temperature displayed or enter a new temperature and press

next

.

Patient Hemoglobin

• The

Patient Hemoglobin (manual entry) feature

allows entry of a hemoglobin value to be used
in the BEb calculation when running a blood
gas test. Press

next

to accept the hemoglobin

value displayed or enter a new value and press

next

.

Reference Range

• The

Reference Range feature allows documen-

tation of the reference range group selected.
Select the reference range group and press

next

.

Sample Type

• The

Sample Type feature allows documentation

of the sample type collected. Press

next

to

accept the sample type displayed or enter a new
type and press

next

.

Sample Site

• The

Sample Site feature allows documentation

of the sample site. Press

next

to accept the

sample site displayed, enter a new site, or select
other to enter a one-to-twelve character

free-text sample site. Press

next

.

• If the sample site selected is Brachial, Femoral, or Radial, highlight “right”

or “left” and press next.

FIO

2

• If a blood gas test is being performed, and the

QA User has selected the Oxygen Therapy
setting FIO2 only, an FIO

2

value may be

entered. Press

next

to accept the FIO

2

displayed, or enter a new FIO

2

value. Press

next

.

Advertising