3.5

a. Passed EQC Test:

Manual

a1. Press

view

to view quantitative results

(optional). Press

next

from each screen to

proceed through results for all analytes.

a2. Press

notes

to select QC notes to attach to the QC record (optional).

a3. Press

print

to obtain a printout. (Results may automatically print,

depending on auto-print setup).

a4. Press

done

to return to the QC TEST OPTIONS menu.

Automatic

a5. Results will print (if the printer is configured to print) and the

analyzer will display the main menu or user
ID screen (if required).

a6. If printer is not configured to print, there

will be no print out. The main MENU will
be displayed.

b. Failed EQC Test:

Manual

b1. Press

help

for assistance in resolving the problem that caused the

failure (optional).

b2. Press

notes

to select QC notes to attach to the record (optional).

b3. Press

print

to obtain a printout. (Results may automatically print,

depending on auto-print setup).

b4. Press

done

to return to the QC TEST OPTIONS menu.

Automatic

b5. If printer is configured to print, “EQC failed” will print and the “EQC

Results failed” screen will display.

b6. Press

done

to return to the main MENU or the User ID entry screen (if

required). Another auto-EQC test will initiate following the next
inactivity period, or a manual EQC test can be performed.