F.4

Figure F-3

c. With the index finger of the other

hand, dispense the sample by

quickly depressing the plunger
straight down in one smooth
motion until it stops

(Figure F-3).

d. Verify that there are no bubbles

or calibrant present in the sample
path, and press

test

. If bubbles or

calibrant are present, press

cancel

to stop the test, discard the single-use

cartridge, and begin again with a new cartridge. Proper injection
technique will prevent air bubbles or calibrant from being present in
the sample path following injection.

7. Following test completion, dispose of the device and cartridge in

accordance with established guidelines for your facility. Do not reuse the
plunger.