IRMA TRUPOINT CARTRIDGES

Each IRMA TruPoint cartridge contains a sensor array and self-contained
calibrant. One patient or liquid QC test is performed on each cartridge.

Figure 1-5

A. Cartridge Leads: electronically connect the cartridge to the analyzer.

B. Luer Injection Port: where the sample collection device attaches to

the cartridge.

C. Sensors: measure analyte concentrations.

D. Calibrant Gel: covers the sensors and is used to calibrate the sensors.

E. Temperature Monitoring Site (Internal): where IRMA TruPoint IR

probe measures and controls sample temperature.

F. Enzyme Pad: may be present on cartridges that employ enzymatic

methodologies.

G. Overflow Indicator: alerts user that the maximum sample volume

(5 mL) that the cartridge can hold has been reached.

H. Waste Reservoir: holds a maximum sample volume of 5 mL.

I. Air Vent: located on the bottom, left side of the waste reservoir.

J. Cal Cap: contains calibrant for cartridges with enzymatic sensors.