6.2

Perform a conditioning charge monthly, or more frequently as the
schedule allows.

A conditioning charge will help ensure optimal battery performance. If a
battery must be removed from the charger before it is completely charged,
the battery should go through a conditioning charge to restore optimal
performance.

Note:

• NiMH batteries (PN 448700) must be charged in a NiMH charger (PN

442900). They are not compatible with the old nickel cadmium (NiCad)
charger (PN 423200).


Charge Times and Test Yields

• NIMH batteries take 5.5-10 hours to fully charge, and yield approximately

30-40 tests.

Battery Life

The rechargeable batteries should retain expected performance characteris-tics
for 500 charge cycles.

Touchscreen Battery Icon

A battery icon displays in the upper-right corner of all screens except the
patient test and QC test screens. A dark meter bar in the battery icon
represents the battery capacity.

Battery (NiMH) at 100 % capacity.

Battery at 50 % capacity.

Battery at 0 % capacity. A flashing battery icon with no
meter bar present represents an empty battery. The
battery should be replaced with a fully charged battery.

AC Adapter in use.

If the battery icon disappears, the battery should be put through a conditioning
charge. See Section 6-Conditioning the Battery.