DT220 Users Manual version 1.a Configuration Parameters

14

Configure Parameters

Press the

Shift key and then the CFG/Test key (Function Key 9) to access the

Configuration Parameters. The following menu is displayed:

1 Scale Parameters

2 Calibrate

3 Com & Printer Ports

4 Event Monitor

5 ID Database

6 PagePrinter

7 Registers

8 Meter Select

9 Passwords

10 System Initialize

Use the

Review and Cont. keys or the numeric keypad to move the pointer to a

configuration menu item and then press the

Enter key.

Press the Clear Entry key to exit configuration mode. The display prompts:

Press P to program changes into ROM

Press X to discard changes

Press P or X to save or discard changes.

Scale Parameters

The scale parameters configure DT220 for the weigh platform or load cell that it is

connected to.

1 A/D speed xxx Filter xx

2 Full Scale xxxxxx

3 Overload Wt xxxxxx

4 Count by: xxxx Alt: xxxx

5 Units Pri: uuuu Alt: uuuu

6 Zero xxx% delay xx band xx

7 Motion delay xx band xx blanks=y/n

8 P.O. Zero = Y

Some of the following parameters are entered in terms of “graduations”. The full-scale

graduations can be found by dividing the full-scale weight by the count-by weight. For

example, if the scale is configured for 500.00lb x 0.05lb, the full-scale graduations is

500.00/0.05 = 10,000 graduations to full scale.
The numbers in parenthesis are the default factory settings.
Note: Jumper TB3 pin 7 to TB6 pin 7 (Function Lock Jumper) to enable scale

configuration and calibration. Jumper enabled as default from the factory RED wire.

Remove to lock configuration after calibration.