DT220 Users Manual version 1.a Configuration Parameters

15

Scale Parameter 1.1: A/D speed (120 samples/second)

The (optional) A/D “Load Cell Module” can operate at speeds from 30 samples/second to

960 samples/second. The default of 120 samples/second bi-polar is optimal for most

applications. Speeds of 480 and 960 are for special applications only (consult factory).

Scale Parameter 1.2: Digital Filter (12)

The digital filter removes scale vibrations and other noise signals from the weight

display. The filter range is between 0 and 20. A setting of 0 provides no filtering; a

setting of 20 provides maximum filtering. The higher the filter number, the longer the

settle time of the weight display. For example:

Filter Setting Settle Time @ 120s/s

3

0.3 sec

6

0.6 sec

9

1.2 sec

12

2.5 sec

15

5.0 sec

Use the smallest filter setting that provides satisfactory display stability.

Scale Parameter 2: Full Scale Graduations (10,000)

Enter the number of graduations at full scale. The full-scale graduations are calculated

using the count-by, decimal position, and full-scale weight.

Full Scale Graduations = Full Scale Weight / count-by.

For example, if Full Scale Weight = 600.0 and count-by = 0.2 then Full Scale

Graduations = 600.0 / 0.2 = 3000.

Scale Parameter 3: Overload Graduations (10,200)

Enter the number of graduations at which the scale is in overload status. The DT220

displays “Error: Over Weight” when the scale weight exceeds overload. For example, if

the Full Scale Graduations is 10,000 grads and the scale is rated for 2% overload, then

Overload Graduations = 10,000 + (10,000 x 0.02) = 10,200.

Scale Parameter 4.1: Count-by (1)

Enter the primary units count-by. For example, if the scale is configured for:

100 kg x 0.02kg then enter 0.02

Scale Parameter 4.2: Alternate Count-by (0.5)

Enter the alternate units count-by.