3-13 Lamp centering

Lamp centering should be done after arc image of lamp is

stablized after main switch was turned ON (5 to 10

minitues later after lamp ON).

In case that frost filter 32FR80 is used, displace it from

light path.

1. Move shutter knob (1) to right (

mark side) to close the

shutter.

2.

Turn front side turret and insert fluorescent mirror unit of B or

IB excitation for visual observation into light path.

(In case that these mirror units are not installed, use other

fluorescent mirror unit. However, in case of fluorescent

mirror unit of U excitation, it is essentially required to observe

through ultraviolet light cutting plate.)

3. Turn rear side turret and insert reference mirror of No.1 into

light path.

4. Move shutter knob (1) to left (

mark side) to open the

shutter.

5. Put objective lens 10x into light path. Put centering target U-

CST on the stage and align center of view with cross line of

U-CST.

6. Turn revolving nosepiece and bring a place where the

objective lens does not exist into light path (objective cap

should also be displaced).

7. Pull out field iris diaphragm lever (2) (to make the view

squeezed).

8. Turn collector lens focusing knob (3) to project arc image on

U-CST. (A)

In case that the image is not projected, turn lamp centering

knob (4).

9.

Turn lamp centering knob (4) and move arc image to center of

right (or left) half view. (B)

10. Insert the Allen screwdriver into mirror focusing screw (5)

located at back panel of lamp house and turns it to bring

mirror arc image in focus. (C)

(In case of BX)

(In case of IX)

TIP

(1)

(1)

(2)

(3)

(4)

(5)