/ Installation of light illuminator

II.

DSU

II .

8 - 3

Page

8-2 Installation of light illuminator

8-2-1 In case of BX

1.

Remove caps of mounting screws (2) on top of light illuminator (1)

with use of tweezers.

2. When using BX61, BX62, BX51 or BX52, put arm lift kit (WI-

ARMAD) on the arm and fix mounting screws (3) (4 pieces) to the

light illuminator firmly, using dedicated

the Allen wrench that

comes with kit.

3. When mounting the light illuminator, put it over slightly forward

position without aligning mounting screws (2) with mounting holes

(4) of microscope body.

4. Next, pressing the light illuminator towards right side and press it

towards rear side so that it would be positioned in correct

mounting position.

5.

While pressing the light illuminator towards rear side, fix mounting

screws (2) (4 pieces) to the light illuminator by turning the

dedicated

the Allen wrench that comes with kit in clockwise

direction.

6.

Attach caps of mounting screws (2) to their original positions.

7. When using BX61, BX62, BX51 or BX52, revolving nosepiece

adapter (DSU-READ) must be attached as follows.

1)

Loosen fixing screw (5) of microscope frame with use of the

dedicated the Allen screwdriver.

2) Align

revolving nosepiece adapter (6) with mount dovetail

from front side of microscope frame and push it and then,

tighten fixing screw (5).

3) Loosen fixing screw (7)

located at right side of revolving

nosepiece adapter, using the Allen screwdriver.

4) Align revolving nosepiece (8) with mount dovetail from front

side of microscope frame and push it and then, tighten fixing

screw (7).