ASSEMBLY

/ Z board setting and installation

II.

DSU

II .

8 - 2 6

Page

Z board installation

You must install Z board after you turn power of control box

to OFF.

1. Loosen fixing knobs (6 pieces) for two slot covers at back side of control

box (1) and remove covers and knobs.

2. Aligning direction of connector of control box and connector of Z board

(2), insert board along with board rails.

Push in until connectors are firmly connected.

3. Using knobs you removed, fix Z board (2). Likewise, attach the other

cover (3).

Keep one piece of slot cover that remained with care as it

may be required for Z board when it is removed.

8-23-3 In case of IX81

For Z board, use type U-ZPCB(T2) for sure.

In case that U-ZPCB(T2) or subsequent is not used, the following

troubles may occur even if correct setting was done.
● Rotation of focusing handle and actual objective lens movement would

be reversed.

● Focusing control may not be able to initialize, depending upon the

position. (Microscope does not boot up).

All DIP switches on Z board, S1, S2 and S3 are turned to OFF at ex-factory.