23-2 in case of bx61w1 – Olympus IX2-DSU BX-DSU User Manual

Page 79

Advertising
background image

ASSEMBLY

/ Z board setting and installation

II.

DSU

II .

8 - 2 5

Page

3. Using knobs you removed, fix Z board (2). Likewise, attach the other

cover (3).

Keep one piece of slot cover that remained with care as it

may be required for Z board when it is removed.

8-23-2 In case of BX61W1

For Z board, use type U-ZPCB(T2) for sure.

In case that U-ZPCB(T2) or subsequent is not used, the following

troubles may occur even if correct setting was done.
● Rotation of focusing handle and actual objective lens movement would

be reversed.

● Focusing control may not be able to initialize, depending upon the

position. (Microscope does not boot up).

All DIP switches on Z board, S1, S2 and S3 are turned to OFF at ex-factory.

Setting change of ONBOARD DIP SW

Use DIP SW set to OFF except the following DIP SW.

Turn DIP SW that is to be changed only to ON .

S2 Changes NO.2 and NO.3 to ON.

S3 Changes NO.2, NO.4 and NO.5 to ON.

If DIP switches are not correctly set, objective lens comes

down and hits the specimen.

Note

Note

TIP

Note

Note

Advertising