22 installation of motor drive board – Olympus IX2-DSU BX-DSU User Manual

Page 77

Advertising
background image

ASSEMBLY

/ DIP switch settings on control box (BX61, BX62 and BX61W1 only)

II.

DSU

II .

8 - 2 3

Page

8-21 DIP switch settings on control box (BX61, BX62 and BX61W1 only)

DIP switch settings should be done after power of control box is

turned OFF.

When you use BX61, BX62 or BX61W1, you must change DIP switch settings on control

box. Set in accordance with the following procedures.

1.

Peel off the seal in front of SW1.

2.

Change DIP switch 7 to ON.

3.

Attach front seal to SW1.

8-22 Installation of motor drive board

You must install motor drive board after you turn power of control

box to OFF.

1.

Loosen four fixing knobs for two option slot covers at left side of back panel

of control box (1) and remove covers and knobs.

2. Aligning

direction

of

connector inside control box with connector (2), insert

the board along with board rails.

Push the board until connectors of motor drive board are inserted into

connectors of mother board inside the control box.

3.

Using knobs you removed, fix motor drive board (2).

Likewise, attach the other cover (3).

Slot cover section must overlap. Attach cover from left side,

seen from back side.

Keep one piece of slot cover that remained with care as it

may be required for motor drive board when it is removed.

Note

(1)

(2)

(3)

Note

Note

Note