ASSEMBLY

/ Installation of mercury lamp

II.

DSU

II .

8 - 7

Page

8-7 Installation of mercury lamp

1.

Loosen socket fixing screw (1), using the Allen screwdriver.

2. Hold upper part of lamp house and pull it vertically to detach socket

section.

If you hold centering knob (2), it may cause a failure.

3.

Place socket section upside down as illustrated.

At ex-factory, a support is provided for transportation. When

changing lamp, old lamp is yet installed. When installing lamp

newly or changing lamp, loosen fixing screws (3) (2 places) that

hold the support or old lamp.

4. Install mercury lamp designated to mounting socket firmly, observing +

polarity to upper base and – polarity to lower base.

Lamp to use should be USH102D (made by Ushio Electric)

or HBO103W/2 (made by OSRAM).

Care must be taken so as not to attach finger mark or dirt

on lamp. If the surface of lamp is contaminated, it may burst

due to distotion of glass. If contaminated, use a cloth like

gauze dipped in mixture of alcohol (3): ether (7) and wipe

out the surface of lamp carefully.

5. Return socket section that the lamp is attached to original position and

tighten fixing screw (1).

Socket section must be lowered vertically, aligning lamp

house outer side and socket section outer side.

(1)

(2)

(3)

(4)

Note

TIP

Note

Note