Change lot, Change lot -85 – Luminex 100 IS User Manual, Version 2.3 User Manual

Page 127

Advertising
background image

x

MAP Technology

Using Luminex 100 IS 2.3 Software

PN 89-00002-00-071 Rev. C

5 - 85

3. The appropriate Standard or Control dialog box opens. To

invalidate or validate all tests click Yes. For only the single test
click No. When invalidating, the Name box turns red with an
asterisk proceeding it. When validating, the Name box reverts to
black text.

4. Click Recalc to recalculate the results if the Auto option was not

selected.

Change Lot

Use the Change Lot command to edit the lot that is applied to the
batch currently opened in data analysis.

‹

To change the lot to another available lot:

1. Click Change Lot (Alt + F8), located at the bottom of Analysis

window, to display the Choose Lot dialog box. See Figure 5-55.

2. The dialog box displays a list of available standard and control

lots that you can apply to a batch. Highlight the desired lot and
click OK to apply the selected lot to the batch opened in data
analysis. Figure 5-55 shows standard and control. Figure 5-56
shows standard only.

Figure 5-55. Choose Lot Dialog Box—Standard and Control