Replay a batch, Replay a batch -56 delete a batch – Luminex 100 IS User Manual, Version 2.3 User Manual

Page 98

Advertising
background image

Luminex 100 IS User Manual Version 2.3

x

MAP Technology

5 - 56

PN 89-00002-00-071 Rev. C

2. Click Clear in the dialog box.

3. Click Yes to confirm that you want to clear the batch.

Replay a Batch

You can reprocess batches through the system multiple times using
Replay Batch. Replay Batch uses the data stored in the run files from
the initial acquisition to reprocess a batch, creating a new batch
output file.

Each time you reprocess a batch using Replay Batch, the system
handles it as if it is a new batch; thus, creating a separate processed
batch entry and output file. The initial batch data and output file
always remain intact and unchanged.

You reprocess a batch using Replay Batch to:

Run as demonstrations to see how the system processes samples
and analyzes the results.

Test a batch using different parameters, such as setting a different
number of events to be collected or using a different bead map or
new formula for analysis, also using a different template.The
number of beads for collection must be less than or equal to the
number of previously collected in the original sample.

If you reprocess a batch with the same template parameters in a
different template, the system obtains results identical to the original
batch. If you reprocess a batch using changed parameters, the system
may obtain different results. When you replay a batch it labels
unknown samples as Pa1, Pa2, Pa3, and so on. If you replay a batch
containing replicates, replicate averaging will not be calculated in
data analysis.

A number of variables can affect the final test results. You may also
change the standards or controls processed with the batch or multi-
batch. These variables may effect your test results:

minimum number of events for acquisition

formula used to analyze the MFI values

standards or controls validation or invalidation

type of analysis (qualitative, quantitative, acquisition only, or
maintenance)

‹

To reprocess samples using Replay Batch:

1. On the Acq. Detail tab, click Replay Batch. The Browse for

Folder dialog box opens displaying the My Batches folder.