Luminex 100 analyzer ventilation filter, Luminex 100 analyzer ventilation filter -9 – Luminex 100 IS User Manual, Version 2.3 User Manual

Page 149

Advertising
background image

x

MAP Technology

Maintenance and Cleaning

PN 89-00002-00-071 Rev. C

6 - 9

11. Hand-tighten the screw on the syringe arm.

12. Plug in the power cord and turn the Luminex 100 analyzer power

on.

13. Prime the system twice, watching for any leaks in the syringe

area.

14. When the prime finishes, close the center door.

Luminex 100 Analyzer
Ventilation Filter

‹

To clean the Luminex 100 analyzer ventilation filter:

1. Disconnect the Luminex 100 analyzer from ac power by turning

off the power switch on the rear of the analyzer, then unplug the
analyzer power cord from the wall source.

2. While facing the Luminex 100 analyzer, push your index finger

up under the right side of the analyzer (in the space between the
Luminex 100 analyzer and the Luminex XYP instrument). When
you feel the filter, push the filter toward the left of the analyzer.
See Figure 6-6.

3. Remove the filter from the left side of the Luminex 100 analyzer.

Figure 6-6 Luminex 100 Analyzer Ventilation Filter

4. Clean the filter with a vacuum or with distilled water. Stand the

filter upright to air dry.

5. Re-install it with the arrows facing up. The filter should click

into place.

6. Plug in and turn on the Luminex 100 analyzer power.