4.3 Systems

761 Compact IC

91

Suppressor

Suppressor

module:

Autostep with

Automatic switching to the next position each

time the injection valve is switched either to the

Fill

or to the

Inject

position.

Use peristaltic pump

If this option is disabled, the peristaltic pump is

not switched on automatically at the start of a

determination or with

Startup hardware

.

Polarity

Selection of the polarity of the output signal:

+

positive polarity (for anions)

negative polarity (for cations)

Measure channel

Display of the data source selected in the

Data

source

window (see section 4.3.5).

Hardware settings

Selection of the

Hardware

menu item using the right mouse button

opens the hardware settings window consisting of the three tabs

Hard-

ware

,

Outputs

and

Alarm stops

.

Hardware
This tab of the hardware settings window defines general parameters

which are set automatically at power on of the instrument.

Remote lines after power on

The remote output lines 1

8 are set to this val-

ues after power on or a manual stop with

Shut-

down hardware

.

Selection:

0

,

1

Thermostat

Operating temperature of the conductivity cell.

Selection:

25

,

30

,

35

,

40

,

45 °C

,

off

Thermostatting functions only if the ambient temperature is at least

5

°C lower than the operating temperature. It normally takes

30

60 min after power on until a temperature stability of

±

0.01 °C is

attained.