2 Installation

761 Compact IC

48

6 Condition system

Rinse the system with eluent until the desired stability of the

baseline is reached (normally 30

60 min; if the eluent is

changed, the establishment of the ion exchanger equilibrium

on the separating column can take longer). After this time the

suppressor module is sufficiently conditioned too.

The instrument is now ready for sample determinations using

the selected system.

Pump tubings are consumable material with a lifetime which depends

on the contact pressure. This is why the tubing cartridges should be

raised completely by loosening snap-action lever on the right-hand

side if the pump is to remain switched off for a considerable length of

time (the set contact pressure remains unchanged).