Metrohm 761 Compact IC User Manual
Page 111
4.4 Methods
761 Compact IC
101
Eluent
Tab
Eluent
of passport with eluent information.
Eluent A
Description of eluent composition (max. 256
characters).
Eluent B
Description of eluent composition (max. 256
characters).
Eluent C
Description of eluent composition (max. 256
characters).
Flow
Flow rate of the high-pressure pump in mL/min.
The flow rate is used to recalculate the time axis
into volumetric units. The system startup value
for the flow rate is entered automatically into
this field at the start of a determination with
Start
with inject
.
Pressure
Pressure in MPa. The measured value for the
pressure is entered automatically into this field
at the start of a determination with
Start with in-
ject
.
Temp.
Temperature in °C. Here you can enter the tem-
perature of the column’s thermostat or the am-
bient temperature.
Comment
Tab
Comment
of passport for entry of free-text user comments into the
chromatogram description. Use this feature to enter any additional in-
formation about the chromatogram not included in other sections of the
method.