9 replacing the suppressor – Metrohm 761 Compact IC User Manual

Page 193

Advertising
background image

5.2 Maintenance and servicing

761 Compact IC

183

128

127

126

125

Fig. 22: Assembling the suppressor

125

Screw nut

127

Suppressor rotor (6.2832.000)

126

Connection piece (6.2832.010)

with input and output leads

128

Suppressor holder






5.2.9

Replacing the suppressor

The suppressor in the suppressor block may have to be replaced in the

following cases:

Irremediable loss of suppressor capacity (reduced

phosphate sensitivity and/or strong rise in baseline)

Irremediable blockage of the suppressor (the suppres-

sor can no longer deliver solutions)

Both the 6.2832.000 Suppressor rotor and the 6.2832.010 Connection

piece with the input and output leads can be replaced. To replace these

components proceed as follows (see Fig. 22):

1 Disconnect suppressor from IC system

Disconnect all input and output leads of the suppressor from

IC system and peristaltic pump.

2 Dismantle suppressor

Unscrew nut 125 from suppressor holder 128.

Pull out connection piece 126 and suppressor rotor 127 from

suppressor holder 128 (the connection piece and the rotor

normally stick together).

Loosen connection piece 126 from suppressor rotor 127.

Advertising