4.3 Systems

761 Compact IC

89

List of program instructions
The following program instructions can be added to the time program

on the

Program

subpage:

Instruction Parameter

entry

Meaning

Valve Inject,

Fill

Switch injection valve to "INJECT" or

"FILL" position.

FullScale

50, 250, 1000

µ

S/cm

Set full scale range to the selected

value.

ICPump on,

off

Switch on or off the high-pressure

pump.

Flow

0.2 ... 2.5 mL/min

Set flow rate of the high-pressure

pump to the desired value.

Pmax

0.0 ... 25.0 MPa

Set maximum pressure limit for the

high-pressure pump to the desired

value.

Pmin

0.0 ... 25.0 MPa

Set minimum pressure limit for the

high-pressure pump to the desired

value.

Remote

0, 1, *, p

Set remote output lines 1...8 to the

desired values. For entry of the first

value, enter

1

,

0

,

p

or

*

. For entry of the

other values, move the cursor in front

of the value to be changed and enter

1

,

0

,

p

or

*

.

Program

END, RESET

The

END

flag can be used to end a

program, especially if the program

time should be longer than the

chromatogram duration. Additional

steps after this flag are not allowed.

The

RESET

flag is used to reset the

parameters to the system startup

values.

Suppressor

Switch

suppressor module to the

next position.

Peristaltic on,

off

Switch on or off the peristaltic

pump.




Remote configuration
On the

Remote configuration

subtab user-defined remote commands can

be defined, which can be inserted into a time program.