5 Notes – Maintenance – Faults

761 Compact IC

170

Storage
Always store the separating columns closed when not in use and filled

in accordance with the manufacturer’s specifications.

Dead volume
Dead volume at the end of a column can be the cause of extreme peak

broadening or splitting (appearance of double peaks). Filling the col-

umn with glass beads (

100

µ

m) frequently improves the separation

efficiency.

Regeneration
If the separation properties of the column have deteriorated, it can be

regenerated in accordance with the column manufacturer’s specifica-

tions. With the separating columns available from Metrohm (see sec-

tion 6.3.2), the instructions for regeneration can be found on the leaflet

enclosed with every column.

In the case of separating columns with carrier material based on silica,

only solutions with pH 2

7 may be used for regeneration, other-

wise the columns could be damaged.

5.1.2 High-pressure

pump

Pulsation dampener
The MF 6.2620.150 Pulsation dampener belongs to the standard ac-

cessories of the 761 Compact IC; its installation is described in sec-

tion 2.6.2. It is used to reduce interfering pulsation in highly sensitive

measurements and also protects the column material against pressure

shocks caused by the injection.

Maintenance
To protect the pump against foreign particles, we advise you to subject

the eluent to microfiltration (0.45

µ

m filter) and siphon the eluent

through the 6.2821.090 Aspirating filter.
In many cases, an unstable baseline (pulsation, flow fluctuations) can

be traced to contaminated valves or faulty, leaky piston seals.
Contaminated valves are cleaned by rinsing with water, RBS solution

or acetone (see section 5.2.6). When the cleaned valves are reinstalled,

you must ensure that the flow direction is correct.
The replacement of piston seals is described in section 5.2.6.
Salt crystals between the piston and the seal are the cause of abrasive

particles, which can enter the eluent. These lead to contaminated

valves, pressure rise and in extreme cases to scratched pistons. It is

thus essential to ensure that no precipitates can appear (see also

section 5.1.3).