8 suppressor module, 1 general information on suppressor module, 2 preparation of the peristaltic pump – Metrohm 761 Compact IC User Manual

Page 48

Advertising
background image

2 Installation

761 Compact IC

38

2.8 Suppressor

module

2.8.1

General information on suppressor module

The Metrohm Suppressor Module MSM for chemical suppression

installed in the 2.761.0020 Compact IC comprises a total of 3 suppres-

sor units which are in turn used for suppression, regenerated with sulfu-

ric acid and rinsed with water. To record every new chromatogram un-

der comparable conditions, work is normally carried out with freshly

regenerated suppressor. Switching is either automatic together with the

valve switching or manual.

The suppressor units must never be regenerated with H

2

SO

4

in the

same flow direction used for the eluent. You should thus always install

the inlet and outlet capillaries as described in section 2.8.4 according

to the scheme shown in Fig. 18.


For operation of the suppressor module, the two-channel peristaltic

pump built into the 2.761.0020 Compact IC is used which conveys the

regeneration solution (normally 20 mmol/L H

2

SO

4

) and the rinsing so-

lution (normally dist. H

2

O) to the suppressor units (flow rate of 0.5

mL/min).
The three inlets and outlets numbered 1...3 on the suppressor module

each have 2 permanently mounted PTFE capillaries, which must be

connected as described in section 2.8.4 (see Fig. 16 and Fig. 18).
To avoid contamination of the suppressor module by foreign particles

or bacterial growth, the two 6.2821.120 Filter units PEEK (see sec-

tion 2.3.6) supplied with the 2.761.0020 Compact IC must be installed

between the peristaltic pump and the inlet capillaries of the suppressor

module.

The suppressor module must never be switched in the dry state as

there is a danger of blocking. Before every switching operation of the

suppressor module, the three suppressor units must have been

rinsed for at least ½ h with eluent, regeneration and rinsing solution.

2.8.2

Preparation of the peristaltic pump

Before start-up the accessories for the 2-channel peristaltic pump built

into the 761 Compact IC must be mounted according to Fig. 15. Pro-

ceed as follows:

1 Attach pump tubings

Loosen both tubing cartridges 48 mounted above pump drive

52 from the holding clamp 50 by pressing down snap-action

lever 51 and remove from mounting pin 53 (see Fig. 16).

Advertising