Metrohm 761 Compact IC User Manual

Page 96

Advertising
background image

4 Operation

761 Compact IC

86

Flow, mL/min

Startup value for flow rate of the high-pressure

pump.

Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Startup value for maximum pressure limit for

high-pressure pump.

Entry range:

0.0

...

25.0 MPa

Pressure min, MPa

Startup value for minimum pressure limit for

high-pressure pump.

Entry range:

0.0

...

25.0

MPa

Full scale,

µ

S/cm

Startup value for full scale range.

Selection:

50

,

250

,

1000

µ

S/cm

Remote lines

Startup value for remote line settings 1

8.

Selection:

0

,

1

Instrument control for connected system

A double-click on the instrument icon or selection of the

Open

menu

item using the right mouse button opens the system settings window.

For a connected system, the

Control

tab is displayed. It allows manual

control of the 761 Compact IC functions and setting of startup values to

be sent to the instrument. This tab shows the current measurement val-

ues for conductivity and pressure.

Conductivity,

µ

S/cm

Live display of measured conductivity.

Pressure, MPa

Live display of measured pressure.
The color settings for this two fields can be

changed by clicking the fields with the right