10 exchanging the pump tubing – Metrohm 761 Compact IC User Manual
Page 195
5.2 Maintenance and servicing
761 Compact IC
185
5.2.10 Exchanging the pump tubing
The pump tubings used by the peristaltic pump are consumable mate-
rial with a limited lifetime and should be exchanged at regular intervals
(approx. every 4 weeks under continuous use).
The working life of pump tubing depends to a considerable extent on
the contact pressure. This is why the contact pressure must be correctly
set as described in section 2.9.2. If the pump is to remain switched off
for a lengthy period of time the tubing cartridges 48 should be raised
completely by loosening the snap-action lever 51 on the right-hand side
(the set contact pressure remains unchanged).
As the pump is always operated on the same side the 6.1826.060
Pump tubings supplied can be used on both sides. To exchange a
pump tubing proceed as follows:
1 Remove old pump tubing
•
Press contact pressure lever 49 on the tubing cartridge 48
down as far as it will go.
•
Release tubing cartridge 48 from holding clamp 50 by press-
ing down snap-action lever 51 and remove from mounting pin
53 (see Fig. 16).
•
Remove old pump tubing 92 or 93.
2 Insert new pump tubing
•
Insert the new pump tubing 92 or 93 in the tubing cartridge
48 as shown in Fig. 15. The white-yellow stopper 94 must
click into the corresponding holder on the left-hand side of
the tubing cartridge.
•
Place the tubing cartridge 48 on mounting pin 53 and press
down on the right-hand side until snap-action lever 51 clicks
into position on holding clamp 50. Take care that no kinks are
formed in the pump tubing.
3 Set contact pressure
•
Switch on peristaltic pump.
•
Press contact pressure lever 49 upwards until the solution
just starts to be drawn in. Then press contact pressure lever
upwards until it clicks once more to obtain optimal contact
pressure.
•
Switch off peristaltic pump.