5 Notes – Maintenance – Faults

761 Compact IC

182

2 Dismantle suppressor

Unscrew nut 125 from suppressor holder 128.

Pull out connection piece 126 and suppressor rotor 126 from

suppressor holder 128 (the connection piece and the rotor

normally stick together).

Loosen connection piece 126 from suppressor rotor 127.

3 Clean input and output leads

Connect each of the 6 capillary tubings attached to connec-

tion piece 126 to the high-pressure pump one after another,

and pump through ultrapure water.

Check whether solution emerges from connection piece 126.

If one of the input or output leads remains blocked, replace

the connection piece (order number 6.2832.010).

4 Clean suppressor rotor

Clean the sealing surface of suppressor rotor 127 using a

lint-free cloth and ethanol.

5 Insert suppressor rotor

Insert suppressor rotor 127 in suppressor holder 128 in such

a way that the tubing connections at the rear of the rotor fit in

the corresponding openings inside the rotor, and that one of

the three holes in the rotor can be seen from below in one of

the openings of the holder.

If the rotor has been inserted correctly, its sealing surface will

be about 4 mm inside the holder. If this is not the case, bring

the rotor into the correct position from below with the aid of a

sharp object (e.g. a screwdriver).

6 Clean connection piece

Clean the sealing surface of connection piece 126 using a

lint-free cloth and ethanol.

7 Insert connection piece

Insert connection piece 126 in suppressor holder 128 in such

a way that connection "1" is at the top, and that the three lugs

on the connection piece fit in the corresponding openings of

the holder.

Screw nut 125 onto the thread of suppressor holder 128

manually (do not use tools).

8 Connect and condition the suppressor

Reconnect the suppressor to the IC system.

Before switching the suppressor to the next position for the

first time, rinse all 3 suppressor units with solution for 5 min.