1.1 Instrument description

761 Compact IC

1

1 Introduction

1.1 Instrument

description

The 761 Compact IC is a PC-controlled system for ion chroma-

tographic analyses. The two following versions are available:

2.761.0010 Compact IC without suppressor module

2.761.0020 Compact IC with suppressor module

The extremely compact housing of the 761 Compact IC contains every-

thing needed to carry out ion chromatography at the highest quality

level:

Injection valve – for individual injections or for use with a sample

changer such as the Metrohm 766 IC Sample Processor

High-pressure pump – extremely low-pulsation double piston

pump with a flow range from 0.2

2.5 mL/min and a maximum

pressure of 25 MPa (250 bar)

Pulsation dampener – even with low-level pressure variations the

pulsation dampener reliably protects the column against damage

Column chamber – the perfect insulation of the housing provides

not only thermally stable conditions for the separation column but

also shields the system against electromagnetic interference

Columns – whether anion columns with or without suppression,

separation columns for cations or organic acids – the 761 Compact

IC can accommodate them all

Suppressor – the Metrohm Suppressor Module (MSM) is already

integrated in the 2.761.0020 Compact IC – pressure-resistant, with

fully automatic regeneration, highest performance and optimal re-

producibility

Peristaltic pump – integrated two-channel peristaltic pump with a

flow rate of 0.5

0.6 mL/min for regeneration and rinsing of the

suppressor module built into the 2.761.0020 Compact IC

Detector – conductivity detector with outstanding temperature

stability. The detector temperature varies by less than 0.01°C and

can be optimally adapted to the ambient conditions.

All components which come into contact with eluent and sample are

metal-free.