6.3 Monitoring The Patient

After the ECG monitoring electrodes are connected, the defibrillator displays a non-diagnostic ECG of the patient’s
heart rhythm and heart rate in a Lead II configuration. While in ECG Monitor Mode, the defibrillator’s shock capability
is disabled; however, the defibrillator continues to evaluate the patient’s ECG for a potentially shockable rhythm.

Monitoring

Monitoring

Monitoring

Plug in Rescue Pads

ECG Monitor Mode Screen

Shockable rhythm detected

If a shockable rhythm is detected, the defibrillator prompts Plug in Rescue Pads.

• Confirm the patient’s condition: Not responsive? Not breathing? No signs of circulation?
• Remove the ECG Monitoring Adapter and connect the defibrillation pads (Rescue Pads) to the AED.
• Apply the defibrillation pads to the patient’s chest, keeping them at least 2.5 cm (one inch) away from the

ECG monitoring electrodes. If necessary, remove the ECG monitoring electrodes.

• Follow the AED’s voice and onscreen text prompts.

31