1. Intr

oduction

A. Speaker. The speaker projects the voice

prompts when the DDU-2000 Series AED is
on. The speaker also emits a “beep” when
the unit is off and has detected a condition
that requires attention from the user or
needs servicing.

B. SHOCK Button. This button will flash when a

shock is recommended. Pressing this button
will deliver a shock when the button is flashing.
This button is disabled at all other times.

C. Display Screen. Color display panel used to

display text and video prompts, messages,
indicators for rescue, unit status, and
maintenance operations. The display screen
provides visual prompts, including CPR
coaching, to assist rescuers with step-by-step
instruction. DDU-2400/2450 models can also
show an ECG trace.

D. ON/OFF Button. This button is used to turn

the DDU-2000 Series AED on and off.

E.

Pads Connector Socket. The pads connector
(item O) is inserted into this socket.

F. Active Status Indicator (ASI). The ASI

indicates the current status of the AED.
This indicator flashes green to indicate
the unit is ready for use and flashes red to
indicate unit needs attention from the user
or needs servicing.

G. Softkey Buttons. Three context sensitive

softkey buttons are used to navigate menus
or select actions.

H. USB Port. The USB port is provided to

perform data recovery and maintenance.
Not to be used during rescue operation.

I. Defibtech Data Card (DDC card).

This optional plug-in card provides enhanced
storage capabilities to the AED.

J. USB and Defibtech Data Card (DDC card)

Access Door. Behind the access door is the
USB connector port and Defibtech Data Card
(DDC card) slot.

K. Unit Serial Number. The unit’s serial

number can be found on the back of the
AED, above the battery pack opening.

L. Battery Pack Opening. This opening is where

the battery pack is inserted into the unit.

M. Battery Pack Eject Release Latch.

This release latch releases the battery pack
from the DDU-2000 Series AED.

N. Pad Storage Area. The pad storage area is

found on the back of the AED allowing the
pads to be stored in a pre-connected state
for rapid deployment during an emergency.

O. Pads Connector. This connector attaches the

patient pads to the unit at the pads connector
socket (item E).

P. Defibrillation Pads. The defibrillation

pads are pads that are placed on the patient.
The pads may be stored in the pad storage
area (item N) on the back of the unit.

Q. Defibrillation Pads Expiration Date

(back side). The defibrillation pads expiration
date is located on the back side of the pads
package. Do not use the pads after the
printed date has passed.

R. Battery Pack Serial Number.

The battery pack’s serial number is located
on the label on the battery pack.

S. Battery Pack. The battery pack provides

a replaceable main power source for the
DDU-2000 Series AED.

T. Battery Pack Expiration Date.

The battery pack expiration date is printed
on the label on the battery pack. Do not
use the battery pack after the printed
date has passed.

U. ECG Monitoring Adapter.

This optional adapter for the DDU-2400 and
the DDU-2450 provides a non-diagnostic
ECG display of the patient’s heart rhythm
for attended patient monitoring.

1.2 The Defibtech DDU-2000 Series AED