Energy – Manual Mode (DDP-2001 Adult and DDP-2003 Non-attenuating Pediatric Defibrillation Pads) (Nominal)

Load Impedance (Ohms)

Selected

Energy

25

50

75

100

125

150

175

25 J

25

25

26

26

26

25

25

50 J

50

50

51

51

51

50

49

70 J

70

70

71

71

72

70

68

100 J

99

100

101

101

102

100

97

150 J

153

151

152

151

153

146

142

200 J

194

195

198

197

201

195

189

DDP-2003 pediatric defibrillation pads Selected Energy range is limited from 25 to 100 J.

Phase Time and Energy (DDP-2002 Attenuating Pediatric Defibrillation Pads)

Patient Impedance

(Ohms)

Phase A Duration

(ms)

Phase B Duration

(ms)

Energy Delivered

(Joules)

25

4.1

4.1

35

50

5.8

3.8

47

75

5.8

3.8

51

100

7.2

4.8

53

125

7.2

4.8

52

150

9

6

53

175

9

6

51

Applies to both AED and Manual Modes.
Note: If impedance is out of range for proper analysis and shock delivery, voice and/or visual prompts will inform the user.