11.2 Battery Packs

Use only Defibtech battery packs in the DDU-2000 Series AED.

DBP-2003 and DBP-2013 Battery Packs

Category

Specification

Model number

DBP-2003
DBP-2013 (Aviation;TSO C-142a)

Main battery type

12VDC, 2800 mAh, Lithium/Manganese Dioxide.
Disposable, recyclable, non-rechargeable.

Capacity

125 shocks or 8 hours of continuous operation.*

Charge time

AED Mode:
• 4 seconds or less (from shock advised)*

Manual Mode:
• 9 seconds or less (150 J)**
• 12 seconds or less (200 J)**

Standby-life (installed in unit)

4 years*

*Typical, new battery, at 25°C

**Typical, new battery depleted by 6 shocks, at 25°C

11.3 Self-Adhesive Defibrillation Pads

Use only Defibtech defibrillation pads with your DDU-2000 Series AED. Defibtech self-adhesive defibrillation pads
have the following characteristics:

Model number

DDP-2001

DDP-2002 and DDP-2003

Type

Adult

Child/Infant < 8 years
< 55 lbs. (25 kg)

Intended use

Disposable

Disposable

Adhesion

Self-adhesive

Self-adhesive

Active gel surface area

77 cm

2

each (nominal)

50 cm

2

each (nominal)

Cable/connector type

Integrated

Integrated

Cable length

122 cm (typical)

122 cm (typical)

Expiration date

2.5 years from date of manufacture

2.5 years from date of manufacture

Note: In the event of a suspected pad defect, the pads should be clearly marked “Not for Use” and returned to
Defibtech, LLC for analysis. (Refer to

Chapter 14

of this manual for contact information regarding returns.)

11.4 ECG Monitoring Adapter / Cable (Optional)

For use only with DDU-2400 and DDU-2450 AEDs.

Model number

DAC-2020 and DAC-2021

Patient connection

Type CF, fully defibrillation-protected

Cable length

2 meters

Patient lead wire designation

Lead type

DAC-2020 (AHA)

DAC-2021 (IEC)

Placement

Positive

Red - LL

Green - F

Left leg

Negative

White - RA

Red - R

Right arm

Reference

Black - LA

Yellow - L

Left arm

Typical lead connection

Lead II

Performance with
DDU-2400/2450 AED

Meets

environmental specifications

for DDU-2000 Series.