Electr omagnetic conf ormity – Defibtech DDU-2450 Series User Manual User Manual

Page 61

Advertising
background image

61

Immunity test

IEC 60601
test level

Compliance
level

Electromagnetic environment –
guidance

Radiated RF
IEC 61000-4-3

10 V/m
80 MHz to
2.5 GHz

80% 5Hz AM
Modulation

10 V/m

Portable and mobile RF
communications equipment should
be used no closer to any part of the
DDU-2000 Series, including cables,
than necessary. The recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter is shown
in the following table.

Interference may occur in
the vicinity of equipment
marked with this symbol.

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.

The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz;
13,553 MHz to 13,567; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the DDU-2000
Series is used exceeds the applicable RF compliance level above, the DDU-2000 Series should be observed to
verify normal operation. If abnormal performance is observed additional measures may be necessary, such as
reorienting or relocating the DDU-2000 Series AED.

Separation Distances

DDU-2000 Series AEDs are intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of a DDU-2000 Series AED can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
DDU-2000 Series AED as recommended below, according to the maximum output of the communications equipment.

Recommended separation distances between portable and mobile

RF communications equipment and DDU-2000 Series AEDs

Separation distance according to frequency of transmitter (m)

Rated maximum
output power of

transmitter (W)

80 MHz to 800 MHz

d = 1.2√P

800 MHz to 2.5 GHz

d = 2.3√P

0.01

0.12

0.23

0.1

0.38

0.73

1

1.20

2.30

10

3.79

7.27

100

12.00

23.00

Electromagnetic Immunity (continued)

1

2. Electr

omagnetic

Conf

ormity

DAC-E2510EN-BB

61

Advertising