Environmental

Category

Specification

Operating/
Maintenance

Temperature

0 – 50°C (32 – 122°F)

Humidity

5% – 95% (non-condensing)

Air Pressure

700 to 1060 hPa (21 to 31 inHg)

Standby/Storage/
Transport

Temperature

0 – 50°C (32 – 122°F)

Humidity

5% – 95% (non-condensing)

Air Pressure

500 to 1060 hPa (15 to 31 inHg)

Altitude

-150 to 4500 meters (-500 to 15,000 feet) per
MIL-STD-810F 500.4 Procedure II

Shock/Drop Abuse Tolerance

MIL-STD-810F 516.5 Procedure IV 48 in, (1.2 meters), any edge,
corner, or surface, in standby mode

Crush test

1,000 lbs (450 kg)

Vibration

MIL-STD-810F 514.5 Category 20 (Ground)
RTCA/DO-160D, Section 8.8.2, Cat R, Zone 2, Curve G (Helicopter)
RTCA/DO-160D, Section 8, Cat H, Zone 2, Curves B & R (Jet Aircraft)

Sealing/Water Resistance

IEC 60529 class IP55; Dust Protected, Protected against water jets
(Battery pack installed)

ESD and EMI (radiated and immunity)

Refer to

Chapter 12

for details

Radio Frequency Emissions Applicable
Directive and Standards

R&TTE Directive 1999/5/EC
ETSI EN 300 220-2 V2.1.2 (2007-06)
ERC RECOMMENDATION 70-03
ETSI EN 301 489-3 V1.4.1 (2002-08)

Aviation

Meets RTCA/DO-160G, Section 21, RF Radiated Emissions,
Category M

Patient Analysis System

The DDU-2000 Series Patient Analysis System ensures that the pad/patient impedance is within the proper range
and analyzes the patient’s ECG rhythm to determine whether a shock is required. On detection of a non-shockable
rhythm, the user is prompted to perform CPR. For shockable rhythms, the AED automatically charges in preparation
for shock delivery.

The patient analysis system will detect electrical “noise” or artifact in the ECG signal that may interfere with accurate
rhythm analysis. This artifact may be caused by excessive motion to the patient or by external electrical noise. When
this artifact is present, the AED will prompt the user to “Stop motion” or “Stop Interference” until the ECG signal is
free of noise and then proceed to analysis.