11 . tec hnical specifications – Defibtech DDU-2450 Series User Manual User Manual

Page 54

Advertising
background image

Waveform Specifications

In AED Mode, the DDU-2000 Series AED delivers a 150 J (Adult) 50 J (Pediatric) Biphasic Truncated Exponential
waveform (AED Mode) to patients with impedances ranging from 25 to 180 ohms. In Manual Mode, the user can
select the energy level.

V

V max

V min

A

B

V1

V1'

t

The waveform is adjusted to compensate for measured patient impedance.
Nominal phase times and energy delivered are shown in the tables below.

Phase Times (DDP-2001 Adult and DDP-2003 Non-attenuating Pediatric Defibrillation Pads)

Patient Impedance

(Ohms)

Phase A Duration

(ms)

Phase B Duration

(ms)

25

2.8

2.8

50

4.1

4.1

75

7.2

4.8

100

9.0

6.0

125

12.0

8.0

150

12.0

8.0

175

12.0

8.0

Energy – AED Mode (DDP-2001 Adult Defibrillation Pads) (Nominal)

Load Impedance (Ohms)

Nominal

25

50

75

100

125

150

175

150 J

153

151

152

151

153

146

142

Energy – AED Mode (DDP-2003 Non-attenuating Pediatric Defibrillation Pads) (Nominal)

Load Impedance (Ohms)

Nominal

25

50

75

100

125

150

175

50 J

50

50

51

51

51

50

49