Aed status screen, 3 entering maintenance mode, Maintenance mode – Defibtech DDU-2450 Series User Manual User Manual

Page 41

Advertising
background image

8.3 Entering Maintenance Mode

Before You Begin: Make certain the DDU-2000 Series AED is turned off and a battery pack is installed.

AED status

OK

Battery status

Expires 01/2018

Pads status

Adult

Expires 06/2018

AED S/N

099999999

Battery S/N

099999999

Software version

V2.2

STEP 1 – Press and release the CENTER softkey button.
Result – The unit will turn on and display the AED Status Screen for a
short period of time.

If the unit does not turn on, check to make sure a good battery pack
is installed. (Refer to the

Troubleshooting

section in Chapter 7 of this

manual.)

STEP 2 – Press the BOTTOM softkey button (to the right of the Tool Icon). Note:

If the bottom softkey button

is not pressed within a short period of time, the unit will automatically turn off.
Result – The unit will enter Maintenance Mode and display the AED Main Menu screen.

8.4 AED Main Menu Screen

The AED Main Menu screen allows the user to view the status of the AED, perform maintenance functions, change
AED options, and access help topics. All maintenance functions are accessed through the AED Main Menu screen.
The user may select from the following options using the softkey buttons:

Rescue now

AED status
AED maintenance
AED options
Rescue options
Help topics
Turn AED off

AED Main Menu

Rescue now – Puts device in Rescue Mode
AED status – Displays current AED status information
AED maintenance – Displays AED Maintenance Menu screen
AED options – Displays AED Options Menu screen
Rescue options – Displays Rescue Options Menu screen
Help topics – Displays Help Topics screen
Turn AED off – Turns unit off

When the user selects “Rescue now,” the unit will exit Maintenance Mode and proceed directly to Rescue Mode.

The other menu options will perform different functions and are described in detail below.

8.5 AED Status Screen

The AED Status screen displays unit-specific data such as current status, battery pack charge state, battery pack
expiration date, defibrillation pads expiration date, unit serial number, battery pack serial number, and software
version number.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to AED status:

AED Main Menu Ú AED status

Note: When the unit is turned off, the AED Status screen can also be accessed by pressing the center softkey button.
What it does: The unit will display the AED Status screen. This is an informational screen only; no action is taken
by the AED.
To exit: To exit the AED Status screen, press and release the BOTTOM softkey button.
The unit will exit the AED Status screen and return to the AED Main Menu screen.

8. Maintenance Mode

41

DAC-E2510EN-BB

Advertising
This manual is related to the following products: