8.6 AED Maintenance Screen

The AED Maintenance screen allows the user to select such options as AED tests, software upgrades, data backups,
and data card functions.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to AED Maintenance:

AED Main Menu Ú

AED maintenance

What it does: The unit will display the AED Maintenance menu screen. This screen will allow the user to navigate
further to perform various maintenance tasks:

• Perform AED test
• Upgrade AED
• Transfer data to card
• Format data card
• Run application from card

To exit: Use the TOP or BOTTOM softkey buttons to scroll to and highlight the selection “Go to main menu.”
Press the CENTER softkey button. The unit will exit the AED Maintenance screen and return to the AED Main
Menu screen.

Perform AED Test

Perform AED test will initiate a system hardware and software self-test.
Note: Running manually initiated AED tests will consume approximately one shock’s worth of battery life.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to Perform AED test:

AED Main Menu Ú AED maintenance Ú Perform AED test

What it does: When the user selects “Perform AED test” selection and presses the CENTER softkey
button, the unit will begin performing the self-test procedure:

The unit says: “Performing AED test“

The unit displays: Testing AED

The unit will then prompt the user to “Press flashing shock button.” Continue to follow the directions
until the test is complete. Once the AED test is complete the unit will verbally and visually report the
status of the AED. The information will be displayed in a pop-up window. The user must press any
softkey button to confirm the test status and return to the AED Maintenance screen.

If self-test passes: The unit will speak and display: “AED OK”

If self-test fails: The unit will display an error screen with text prompts providing instructions to
address the condition.

Note: If self-test fails, the user should follow the text prompts to address the condition requiring
attention or refer to the “

Troubleshooting

section in Chapter 7 of this manual.

To exit: Press any softkey button. The self-test status pop-up window will clear and the display will
return to the AED Maintenance menu screen.

Upgrade AED

The Upgrade AED menu selection is a method of performing a unit upgrade and will activate the
system upgrade procedure from a Defibtech Data Card (DDC card) containing an upgrade application.
Note: Upgrades can also be performed directly from the AED Status Screen if an upgrade card is
present when the AED Status Screen mode is launched.
Before you begin: Be sure the unit is in Maintenance Mode.

8. Maintenance Mode