7 aed options screen, Maintenance mode – Defibtech DDU-2450 Series User Manual User Manual

Page 44

Advertising
background image

To enter: Navigate to Format data card:

AED Main Menu Ú AED maintenance Ú Format data card

What it does: The unit will format the DDC card that is inserted in the AED:

The unit says: “Formatting data card”

The unit displays: Formatting Data Card

When unit finishes formatting the DDC card, the unit will return to the menu.

WARNING

Do not turn off the unit or remove the battery
pack or data card until the operation is complete.

Note: If the data card is not inserted, the unit will speak and display “Data Card Missing.” (Refer to the

Installing The Defibtech Data Card (DDC card)

” section in Chapter 3 of this manual.)

To exit: When the unit finishes formatting the data card, it will automatically return to the AED
Maintenance menu screen.

CAUTION

Using non-Defibtech Data Cards (DDC cards)
may damage the unit and will void the warranty.

Run Application From Card

Run application from card will initiate a card application on the Defibtech Data Card (DDC card).
The most typical application is a software upgrade.
Before you begin: Be sure the unit is in Maintenance Mode. Be sure a DDC Card, with a card
application, is installed in the device. (Refer to the

Installing The Defibtech Data Card (DDC card)

section in Chapter 3 of this manual.)
To enter: Navigate to Run application from card:

AED Main Menu Ú AED maintenance Ú Run application from card

WARNING

Do not turn off the unit or remove the battery
pack or data card until the operation is complete.

Note: If a data card is not inserted, the unit will speak and display “Data Card Missing.”
(Refer to the “

Installing The Defibtech Data Card (DDC card)

” section in Chapter 3 of this manual.)

To exit: When the unit finishes performing the application, follow the displayed and spoken
instructions.

8.7 AED Options Screen

To manually configure AED options such as time, date, volume, and audio recording, select the AED options
from the AED Main Menu screen.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to AED options:

AED Main Menu Ú AED options

8. Maintenance Mode