Cleaning and disposal, Troubleshooting tips, Monitoring continuous temperature – Physio-Control LIFEPAK 15 User Manual

Page 108

Advertising
background image

4-64

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring Continuous Temperature

Cleaning and Disposal

Temperature probes are disposable and intended for single-patient use. Do not clean and reuse

temperature probes. Dispose of the contaminated waste according to local protocols.

Temperature adapter cables are reusable and may be cleaned. To clean the reusable temperature

cable:

1. Disconnect the cable from the monitor.

2. Use a clean, soft cloth dampened with a germicidal solution to wipe clean. See "Cleaning the

Device" on page 10-15 for a list of acceptable cleaning solutions.

3. Allow to dry before reconnecting the cable to the monitor.

For information about cleaning the device, see "Cleaning the Device" on page 10-15.

Troubleshooting Tips

Table 4-11 Troubleshooting Tips for Temperature Monitoring

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

CHECK SENSOR

message

appears and value is “---”

Temperature value is out of
range

• Check that probe is positioned

properly.

Temperature probe is
dislodged or positioned
incorrectly

• Check that probe is positioned

properly.

Probe not connected to
cable, or cable not
connected to device

• Check that probe and cable are

connected properly.

Damaged cable or probe

• Replace damaged cable or probe.

CHECK SENSOR

message

appears while value is
displayed

Temperature probe is
dislodged and value is below
31°C (87.8°F)

• Check that probe is positioned

properly.

Temperature probe is
dislodged and value is above
41.0°C (105.8°F)

• Check that probe is positioned

properly.

TEMP: ACCURACY OUTSIDE
LIMITS

message appears

and value is XXX

Temperature accuracy check
failed

• Turn device off and then on again.
• If problem persists, contact

qualified service personnel.

XXX appears in place of
temperature reading

Temperature module is not
calibrated

• Turn device off and then on again.
• If problem persists, contact

qualified service personnel.

Temperature module failed

• Turn device off and then on again.
• If problem persists, contact

qualified service personnel.