Paddles ecg monitoring procedure, Monitoring – Physio-Control LIFEPAK 15 User Manual

Page 51

Advertising
background image

MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-7

2. Place the other therapy electrode or

STERNUM

paddle on the patient’s upper right torso, lateral

to the sternum and below the clavicle as shown in Figure 4-1.

Special Situations for Electrode or Paddle Placement

When placing therapy electrodes or standard paddles, be aware of the special requirements in

the following possible situations:

Obese Patients or Patients with Large Breasts

Apply therapy electrodes or standard paddles to a flat area on the chest, if possible. If skin folds

or breast tissue prevent good adhesion, it may be necessary to spread skin folds apart to create

a flat surface.

Thin Patients

Follow the contour of the ribs and spaces when pressing the therapy electrodes or standard

paddles onto the torso. This limits air spaces or gaps under the electrodes and promotes good

skin contact.

Patients with Implanted Devices Such as Pacemakers or Defibrillators

If possible, place therapy electrodes or standard paddles away from implanted device.

Paddles ECG Monitoring Procedure

To monitor using standard paddles or therapy electrodes:

1. Press

ON

.

2. Prepare the patient’s skin:

• Remove all clothing from the patient’s chest.

• Remove excessive chest hair as much as possible. Avoid nicking or cutting the skin if

using a shaver or razor. If possible, avoid placing electrodes over broken skin.

• Clean and dry the skin, if necessary. Remove any medication patches and ointment on the

patient’s chest.

• Briskly wipe the skin dry with a towel or gauze. This mildly abrades the skin and removes

oils, dirt, and other debris for better electrode adhesion to the skin.

• Do not use alcohol, tincture of benzoin, or antiperspirant to prep the skin.

3. Apply the standard paddles or therapy electrodes in the anterior-lateral position. For therapy

electrodes, confirm that the package is sealed and the Use By date is not passed. For

standard paddles, apply conductive gel over the entire electrode surface.

4. Connect the therapy electrodes to the therapy cable.

5. Select

PADDLES

lead.