Electromagnetic emissions, Essential performance, Electromagnetic compatibility guidance – Physio-Control LIFEPAK 15 User Manual

Page 271: Appendix

Advertising
background image

APPENDIX

D

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

D-1

Electromagnetic Compatibility Guidance

Electromagnetic Emissions

Essential Performance

The LIFEPAK 15 monitor/defibrillator maintains safe and effective performance of the defibrillation

therapy and patient monitoring functions when operated in the electromagnetic environment

specified in Table D-2 through Table D-4.

Table D-1 Guidance and Manufacturer’s Declaration - Electromagnetic Emissions

The LIFEPAK 15 monitor/defibrillator is intended for use in the electromagnetic environment specified
below. The customer or the user of the LIFEPAK 15 monitor/defibrillator should assure that it is used in
such an environment.

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF emissions

CISPR 11

Group 1

The LIFEPAK 15 monitor/defibrillator uses RF energy only for
its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.

RF emissions

CISPR 11

Class B

The LIFEPAK 15 monitor/defibrillator is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.

Harmonic
emissions

IEC 61000-3-2

Not applicable

Voltage
fluctuations/
flicker emissions

IEC 61000-3-3

Not applicable