10-19

AUXILIARY POWER

LED not

illuminated

Power adapter not properly
connected to auxiliary power
source

• Check that power adapter is properly

connected to auxiliary power.

Power adapter not properly
connected to monitor/
defibrillator

• Check that power adapter is properly

connected to monitor/defibrillator.

Defective power adapter or
cables

• Replace with working power adapter

and cables.

BATTERY CHARGING

LED on

monitor/defibrillator not
illuminated

Power adapter not properly
connected to auxiliary power
source or monitor/
defibrillator

• Check that power adapter is

connected properly.

Battery not properly inserted
in battery well

• Check that battery is properly

inserted in battery well.

Unable to charge battery
with power adapter because
battery charge level is too
low.

• Charge battery in Station-Mobile or

REDI-CHARGE battery charger if
available.

• Replace battery.

No batteries installed

• Install at least one battery.

Defective battery

• Remove battery from service and

replace with working battery.

Unrecognized battery

• Only use battery that is approved for

use with the LIFEPAK 15 monitor/
defibrillator.

Incompatible power adapter
connected to the monitor/
defibrillator

• Only use power adapter that is

approved for use with the
LIFEPAK 15 monitor/defibrillator.

Defective power adapter or
cables

• Replace with working power adapter

and cables.

Monitor/defibrillator unable
to recognize installed
battery

• Contact qualified service personnel.

CANNOT CHARGE BATTERY

message appears

Defective battery

• Remove battery from service and

replace with working battery.

Defective power adapter

• Replace with working power adapter

and cables.

Device unable to charge
battery or batteries

• Contact qualified service personnel.

Table 10-2 General Troubleshooting Tips (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION