Monitoring etco2 – Physio-Control LIFEPAK 15 User Manual

Page 96

Advertising
background image

4-52

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring ETCO2

EtCO

2

values are erratic

FilterLine connection to device
is loose

• Twist FilterLine connector

clockwise until tight and
firmly seated.

A leak in the FilterLine set

• Check for connection leaks

and line leaks to patient,
and correct, if necessary.

A mechanically ventilated
patient breathes spontaneously
or patient is talking

• No action required.

EtCO

2

values are consistently

higher than expected

Physiological cause such as
COPD

• None.

Inadequate ventilation

• Check ventilator, increase

ventilatory rate/bagging.

Patient splinting during
breathing

• Supporting measures such

as pain relief.

Improper calibration

• Contact qualified service

personnel.

EtCO

2

values are consistently

lower than expected

FilterLine connection to device
is loose

• Twist FilterLine connector

clockwise until tight and
firmly seated.

Physiological cause

• See Physiological factors in

"CO2 Detection" on
page 4-50.

Hyperventilation

• Check ventilator, decrease

ventilatory rate/bagging.

Improper calibration

• Contact qualified service

personnel.

CO

2

waveform stays elevated

for several seconds

Expiration is prolonged due to
bagging technique

• Release bag reservoir

completely with expiration.
Observe that elevated
baseline returns to normal
level.

Sudden extreme increase in
EtCO

2

Fluid has entered CO

2

module

• Contact qualified service

personnel.

XXX

appears instead of EtCO

2

value

CO

2

module malfunction

• Turn device off and then on

again.

• If problem persists, contact

qualified service personnel.

There is no EtCO

2

value and the

CO

2

waveform is flat

Measured CO

2

is less than

3.5 mmHg

• See "CO2 Detection" on

page 4-50.

Table 4-6 Troubleshooting Tips for EtCO

2

Monitoring (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

Advertising