Monitoring spo2, spco, and spmet – Physio-Control LIFEPAK 15 User Manual

Page 78

Advertising
background image

4-34

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring SpO2, SpCO, and SpMet

SPO2: NO SENSOR DETECTED

message appears

Sensor not connected to patient
or cable disconnected from
monitor/defibrillator

• Check that sensor and cable

are connected properly.

• Check that appropriate

sensor is in use.

Damaged cable or sensor

• Replace damaged cable or

sensor.

No SpO

2

, SpCO, or SpMet

value (---) is displayed

Sensor may be too tight

• Reposition sensor.
• Relocate sensor.

Patient is in cardiac arrest or
shock

• Check patient.

Oximeter may be performing
self-calibration or self-test

• Wait for completion.
• If values do not display

within 30 seconds,
disconnect and reconnect
sensor. If values do not
display within another 30
seconds, replace sensor.

Defibrillator shock just
delivered

• None. If values do not

display within 30 seconds,
disconnect and reconnect
sensor. If values do not
display within another 30
seconds, replace sensor.

High intensity lights (such as
pulsating strobe lights) may be
interfering with performance

• Cover sensor with opaque

material, if necessary.

Damaged cable or sensor

• Replace damaged cable or

sensor.

Different SpCO or SpMet
measurements on same patient

Every measurement, even on
the same patient, can be
different

• Confirm by taking three

measurements: ring finger,
middle finger, and then
index finger; average the
results.

XXX

appears in place of SpO

2

reading

SpO

2

module failed.

Internal cable failed.

• Turn device off and then on

again.

• If problem persists, contact

qualified service personnel.

Table 4-4 Troubleshooting Tips for SpO

2

, SpCO, and SpMet (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

Advertising